FAQ: Apoptosis, Necrosis, and Cell Viability Kits

Apoptosis and Necrosis Kit FAQs

Where can I find the expiration date or shelf life of a product?

Bioscience kits
The guaranteed shelf life from date of receipt for bioscience kits is listed on the product information sheet. Some kits have an expiration date printed on the kit box label, this is the guaranteed shelf life date calculated from the day that the product shipped from our facility. Kits often are functional for significantly longer than the guaranteed shelf life. If you have an older kit in storage that you wish to use, we recommend performing a small scale positive control experiment to confirm that the kit still works for your application before processing a large number of samples or precious samples.

Antibodies and other conjugates
The guaranteed shelf life from date of receipt for antibodies and conjugates is listed on the product information sheet. Antibodies and other conjugates often are functional for significantly longer than the guaranteed shelf life. If you have an older conjugate in storage that you wish to use, we recommend performing a small scale positive control experiment to confirm that the product still works for your application before processing a large number of samples or precious samples.

For lyophilized antibodies, we recommend reconstituting the antibody with glycerol and antimicrobial preservative like sodium azide for the longest shelf life (note that sodium azide is not compatible with HRP-conjugates).

Chemicals, dyes, and gel stains
Biotium guarantees the stability of chemicals, dyes, and gel stains for at least a year from the date you receive the product. However, the majority of these products are highly stable for many years, as long as they are stored as recommended. Storage conditions can be found on the product information sheet or product safety and data sheet, material safety data sheet, and on the product label. Fluorescent compounds should be protected from light for long term storage.

If you have a Biotium compound that has been in storage for longer than one year that you wish to use, we recommend performing a small scale positive control experiment to confirm that the compound still works for your application before processing a large number of samples or precious samples.

Expiration date based on date of manufacture (DOM)
If your institution requires you to document expiration date based on date of manufacture for reagents, please contact techsupport@biotium.com for assistance.

Chemical products with special stability considerations:

Esters

Ester compounds include the following:
• Succinimidyl esters (SE, also known as NHS esters), such as our amine-reactive dyes
• Acetoxymethyl esters (AM esters) such as our membrane-permeable ion indicator dyes
• Diacetate-modified dyes, like ViaFluor™ 405, CFDA, and CFDA-SE cell viability/cell proliferation dyes

Ester dyes are stable in solid form as long as they are protected from light and moisture. Esters are not stable in aqueous solution. Concentrated stock solutions should be prepared in anhydrous DMSO (see Biotium catalog no. 90082). Stock solutions in anhydrous DMSO can be stored desiccated at -20°C for one month or longer. Esters should be diluted in aqueous solution immediately before use. Succinimidyl esters (SE) should be dissolved in a solution that is free of amine-containing compounds like Tris, glycine, or protein, which will react with the SE functional group. AM esters and diacetate compounds should be dissolved in a solution that is free of serum, because serum could contain esterases that would hydrolyze the compound.

A note on CF™ dye succinimidyl ester stability
Succinimidyl esters are generally susceptible to hydrolysis, which can result in lower labeling efficiency. Heavily sulfonated dyes, such as the Alexa Fluor® dyes, DyLight® dyes and IRDyes® are particularly hygroscopic, worsening the hydrolysis problem. For example, the percent of active Alexa Fluor® 488 succinimidyl ester (SE) could be well below 50% by the time of application (according to the manufacturer’s product datasheet). In a number of Alexa Fluor® SE reactive dyes, the SE group is derived from an aromatic carboxylic acid, while in all of Biotium’s CF™ dyes the SE group is prepared from an aliphatic carboxylic acid. This structural difference reduces the susceptibility of CF™ dye SE reactive groups to hydrolysis, resulting in relatively stable reactive dyes with consistently higher labeling efficiency compared to other SE derivatives of other fluorescent dyes.

Maleimides, MTS and thiosulfate dyes
Like the succinimidyl ester dyes, these dyes are also susceptible to hydrolysis, although generally to a much lower degree. Thus, for long term storage, anhydrous DMSO is recommended for making stock solutions.

Other reactive dyes
Amines, aminooxy (also known as oxylamine), hydrazide, azide, alkyne, BCN, and tyramide reactive dyes, as well as dye free acids, are generally stable in aqueous solution when stored at -20°C for 6-12 months or longer, as long as no compounds are present that may react with the dye’s functional group. See the product information sheets for specific reactive dyes more information.

Coelenterazines and D-luciferin

Coelenterazines are stable in solid form when stored as recommended; they are not stable in aqueous solution. Concentrated coelenterazine stock solutions (typically 1-100 mg/mL) should be prepared in ethanol or methanol; do not use DMSO or DMF to dissolve coelenterazines, because these solvents will oxidize the compounds. Ethanol or methanol stocks of coelenterazine can be stored at -20°C or below for six months or longer; alcohol stocks may evaporate during storage, so use tightly sealing screw cap vials and wrap the vials with Parafilm for long term storage. If the solvent evaporates, the coelenterazine will still be present in the vial, so note the volume in the vial prior to storage so that you can adjust the solvent volume to correct for evaporation if needed. Prepare working solutions in aqueous buffers immediately before use. Coelenterazines are stable for up to five hours in aqueous solution.

Aquaphile™ coelenterazines are water soluble formulations of coelenterazines. They are stable in solid form when stored as recommended. Aquaphile™ coelenterazines should be dissolved in aqueous solution immediately before use. They are stable for up to five hours in aqueous solution.

Note that coelenterazines are predominantly yellow solids, but may contain dark red or brown flecks. This does not affect product stability or performance. If your coelenterazine is uniformly brown, then it is oxidized and needs to be replaced.

D-luciferin is stable in solid form and as a concentrated stock solution when stored as recommended; it is not stable at dilute working concentrations in aqueous solution. Prepare concentrated D-luciferin stock solutions (typically 1-100 mg/mL) in water, and store in aliquots at -20°C or below for six months or longer. Prepare working solutions immediately before use.

My product arrived at room temperature, but it’s supposed to be stored at 4°C or -20°C. Is it still ok?

Most of our products are stable at room temperature for many days, but we recommend storage at 4°C or -20°C to prolong shelf life. In the case of many of our aqueous dye solutions, the compounds are very stable at room temperature, but we recommend cold storage to prevent the growth of mold or other microbes over time. Therefore, to save on shipping costs, products with recommended storage at 4°C or -20°C may ship at ambient temperature without affecting product performance. When you receive the product, place it under the recommended storage conditions.

My product was accidentally left out at room temperature or exposed to light. Is it ruined?

Most of our products are stable at room temperature for many days, so in all likelihood the product will still work just fine. To be on the safe side, we recommend performing a small scale positive control experiment to confirm that the product still works for your application before processing a large number of samples or precious samples.

One exception that we are aware of is GelGreen™, which is more sensitive to light exposure than most of our other fluorescent dyes. If GelGreen™ is exposed to ambient light for a prolonged period of time (days to weeks), its color will change from dark orange to brick red. If this occurs, the GelGreen will no longer work for gel staining.

 

Can I use the kits for staining fixed cells or tissues?

No. Fixation, freezing, sectioning, or dissociation of tissues can affect the PS on the outer leaflet and compromise membrane integrity. Both Annexin V and Ethidium Homodimer III rely upon the presence of intact membranes in healthy cells to accurately distinguish healthy cells from apoptotic or necrotic cells. To detect apoptosis in fixed cells and tissues we recommend our TUNEL kits. Currently we are not aware of a fluorescent probe that specifically detects necrotic cells in fixed tissues. Necrotic cells in tissue sections are identified based on morphological criteria.

Can I use the kits for staining live embryos, tissue explants or organotypic slice cultures?

The assays are designed to stain dissociated cells in culture and have not been validated for organ culture. Annexin V staining of early chicken and mammalian embryos in culture has been reported in the scientific literature. For staining of living tissues, the specimen would need to be thin enough to allow exposure of the cells to the 36 kDa Annexin V protein. Also, damage to cell membranes from dissection or sectioning of tissues could result in high background staining.

Can I fix cells after staining?

Yes, cells can be fixed with formaldehyde after staining. Because Annexin V staining is dependent on calcium, all buffers used for washing and fixation should contain 1.25 mM CaCl2. Fixation may increase the background signal from Ethidium Homodimer III. Wash the cells several times to remove unbound Ethidium Homodimer III before fixation.

What species does Annexin V cross-react with?

Annexin V is a 36 kDa protein that binds to the phospholipid phosphatidylserine. Therefore Annexin V binding is not species-specific.

What is the source of your Annexin V?

The Annexin V protein that we use is a recombinant protein made in E. coli.

What is the purpose of the binding buffer?

The binding buffer is an isotonic buffer containing calcium, which is essential for the binding of Annexin V to phosphatidylserine.

I am using fewer cells than suggested in the protocol, should I adjust the amount of CF488A- or FITC-Annexin V and Ethidium Homodimer III in the assay?

No, the concentration of Annexin V and Ethidium Homodimer III should be kept constant regardless of cell number. However, the staining concentrations can be increased or decreased if necessary to optimize staining.

I am afraid that I’m losing some of my apoptotic cells during processing of cells from an adherent culture, how can I make sure that they are accounted for?

You may wish to process floating and detached cells separately – collect the washes and spin them down and then stain those cells using the protocol for suspension cells. Alternatively you may collect floating cells, detach the adherent cells with trypsin (without EDTA), pool the cells together, and use the suspension cell protocol for staining.

What can I do to fix low signal or high background staining?

The concentration of Annexin V and Ethidium Homodimer III can be increased or decreased if necessary to optimize staining for different cell types.


Cell Viability Kit FAQs

Where can I find the expiration date or shelf life of a product?

Bioscience kits
The guaranteed shelf life from date of receipt for bioscience kits is listed on the product information sheet. Some kits have an expiration date printed on the kit box label, this is the guaranteed shelf life date calculated from the day that the product shipped from our facility. Kits often are functional for significantly longer than the guaranteed shelf life. If you have an older kit in storage that you wish to use, we recommend performing a small scale positive control experiment to confirm that the kit still works for your application before processing a large number of samples or precious samples.

Antibodies and other conjugates
The guaranteed shelf life from date of receipt for antibodies and conjugates is listed on the product information sheet. Antibodies and other conjugates often are functional for significantly longer than the guaranteed shelf life. If you have an older conjugate in storage that you wish to use, we recommend performing a small scale positive control experiment to confirm that the product still works for your application before processing a large number of samples or precious samples.

For lyophilized antibodies, we recommend reconstituting the antibody with glycerol and antimicrobial preservative like sodium azide for the longest shelf life (note that sodium azide is not compatible with HRP-conjugates).

Chemicals, dyes, and gel stains
Biotium guarantees the stability of chemicals, dyes, and gel stains for at least a year from the date you receive the product. However, the majority of these products are highly stable for many years, as long as they are stored as recommended. Storage conditions can be found on the product information sheet or product safety and data sheet, material safety data sheet, and on the product label. Fluorescent compounds should be protected from light for long term storage.

If you have a Biotium compound that has been in storage for longer than one year that you wish to use, we recommend performing a small scale positive control experiment to confirm that the compound still works for your application before processing a large number of samples or precious samples.

Expiration date based on date of manufacture (DOM)
If your institution requires you to document expiration date based on date of manufacture for reagents, please contact techsupport@biotium.com for assistance.

Chemical products with special stability considerations:

Esters

Ester compounds include the following:
• Succinimidyl esters (SE, also known as NHS esters), such as our amine-reactive dyes
• Acetoxymethyl esters (AM esters) such as our membrane-permeable ion indicator dyes
• Diacetate-modified dyes, like ViaFluor™ 405, CFDA, and CFDA-SE cell viability/cell proliferation dyes

Ester dyes are stable in solid form as long as they are protected from light and moisture. Esters are not stable in aqueous solution. Concentrated stock solutions should be prepared in anhydrous DMSO (see Biotium catalog no. 90082). Stock solutions in anhydrous DMSO can be stored desiccated at -20°C for one month or longer. Esters should be diluted in aqueous solution immediately before use. Succinimidyl esters (SE) should be dissolved in a solution that is free of amine-containing compounds like Tris, glycine, or protein, which will react with the SE functional group. AM esters and diacetate compounds should be dissolved in a solution that is free of serum, because serum could contain esterases that would hydrolyze the compound.

A note on CF™ dye succinimidyl ester stability
Succinimidyl esters are generally susceptible to hydrolysis, which can result in lower labeling efficiency. Heavily sulfonated dyes, such as the Alexa Fluor® dyes, DyLight® dyes and IRDyes® are particularly hygroscopic, worsening the hydrolysis problem. For example, the percent of active Alexa Fluor® 488 succinimidyl ester (SE) could be well below 50% by the time of application (according to the manufacturer’s product datasheet). In a number of Alexa Fluor® SE reactive dyes, the SE group is derived from an aromatic carboxylic acid, while in all of Biotium’s CF™ dyes the SE group is prepared from an aliphatic carboxylic acid. This structural difference reduces the susceptibility of CF™ dye SE reactive groups to hydrolysis, resulting in relatively stable reactive dyes with consistently higher labeling efficiency compared to other SE derivatives of other fluorescent dyes.

Maleimides, MTS and thiosulfate dyes
Like the succinimidyl ester dyes, these dyes are also susceptible to hydrolysis, although generally to a much lower degree. Thus, for long term storage, anhydrous DMSO is recommended for making stock solutions.

Other reactive dyes
Amines, aminooxy (also known as oxylamine), hydrazide, azide, alkyne, BCN, and tyramide reactive dyes, as well as dye free acids, are generally stable in aqueous solution when stored at -20°C for 6-12 months or longer, as long as no compounds are present that may react with the dye’s functional group. See the product information sheets for specific reactive dyes more information.

Coelenterazines and D-luciferin

Coelenterazines are stable in solid form when stored as recommended; they are not stable in aqueous solution. Concentrated coelenterazine stock solutions (typically 1-100 mg/mL) should be prepared in ethanol or methanol; do not use DMSO or DMF to dissolve coelenterazines, because these solvents will oxidize the compounds. Ethanol or methanol stocks of coelenterazine can be stored at -20°C or below for six months or longer; alcohol stocks may evaporate during storage, so use tightly sealing screw cap vials and wrap the vials with Parafilm for long term storage. If the solvent evaporates, the coelenterazine will still be present in the vial, so note the volume in the vial prior to storage so that you can adjust the solvent volume to correct for evaporation if needed. Prepare working solutions in aqueous buffers immediately before use. Coelenterazines are stable for up to five hours in aqueous solution.

Aquaphile™ coelenterazines are water soluble formulations of coelenterazines. They are stable in solid form when stored as recommended. Aquaphile™ coelenterazines should be dissolved in aqueous solution immediately before use. They are stable for up to five hours in aqueous solution.

Note that coelenterazines are predominantly yellow solids, but may contain dark red or brown flecks. This does not affect product stability or performance. If your coelenterazine is uniformly brown, then it is oxidized and needs to be replaced.

D-luciferin is stable in solid form and as a concentrated stock solution when stored as recommended; it is not stable at dilute working concentrations in aqueous solution. Prepare concentrated D-luciferin stock solutions (typically 1-100 mg/mL) in water, and store in aliquots at -20°C or below for six months or longer. Prepare working solutions immediately before use.

My product arrived at room temperature, but it’s supposed to be stored at 4°C or -20°C. Is it still ok?

Most of our products are stable at room temperature for many days, but we recommend storage at 4°C or -20°C to prolong shelf life. In the case of many of our aqueous dye solutions, the compounds are very stable at room temperature, but we recommend cold storage to prevent the growth of mold or other microbes over time. Therefore, to save on shipping costs, products with recommended storage at 4°C or -20°C may ship at ambient temperature without affecting product performance. When you receive the product, place it under the recommended storage conditions.

My product was accidentally left out at room temperature or exposed to light. Is it ruined?

Most of our products are stable at room temperature for many days, so in all likelihood the product will still work just fine. To be on the safe side, we recommend performing a small scale positive control experiment to confirm that the product still works for your application before processing a large number of samples or precious samples.

One exception that we are aware of is GelGreen™, which is more sensitive to light exposure than most of our other fluorescent dyes. If GelGreen™ is exposed to ambient light for a prolonged period of time (days to weeks), its color will change from dark orange to brick red. If this occurs, the GelGreen will no longer work for gel staining.

 

What is the difference between MTT, XTT and resazurin?

MTT and XTT are colorimetric based assays, while resazurin can be measured using colorimetric or fluorescence detection. MTT is not a soluble product, so the cells must be lysed to solubilize the formazan salt before absorbance can be measured. XTT and resazurin do not require cell lysis, allowing kinetic monitoring of the same samples at different timepoints.

What if the signal in my negative control sample seems to be high?

Serum and albumin in medium may lead to false positives. It is best to use medium containing less than 10% added protein.

I can see a difference in the density of my cells, why doesn’t the assay reflect the difference?

Check to make sure your cell densities are in the linear range of the assay. Too few cells may fall below the limit of detection, and too many cells may saturate the OD. Perform an initial experiment to check the best linear range for your cell type and experimental set-up. You may need to vary the cell density assay incubation time. The kit protocols provide general guidelines and may need to be optimized for your experimental system.

Why does my sample with the most dense cells seem have the least amount of resazurin fluorescence?

Resazurin is reduced by cells to the fluorescent product resorufin. Resorufin can be reduced further to a non-fluorescent compound. Therefore the densest wells may have lowest fluorescence due to over-reduction of the substrate. Please see the product information sheet for more details. The kit protocol provides general guidelines and may need to be optimized empirically for your experimental system. You may need to vary cell density or assay incubation time to ensure that your samples fall in the linear range of the kit.

What is the difference between resazurin and alamarBlue®?

AlamarBlue® contains resazurin and additional compounds to prevent the over-reduction of resazurin to a non-fluorescent product. These additives also slow the rate of generation of the fluorescent product. Consequently, the alamarBlue® assay requires longer incubation times compared to resazurin.

What is the purpose of the background wavelength? Do I have to use it?

It is recommended to measure the background wavelength; however, the background should be very low compared to the signal. Measuring the background confirms that there is no unexpected signal (for example, from the plate itself) that might interfere with the reading.

Do I have to use the OD specified in the product information sheet?

No, the absorbance spectrum is broad, but the wavelength used for detection should be close to the recommended wavelength.